1.
Chinese Journal of Radiological Health ; (6): 407-411, 2021.
Article in Chinese | WPRIM | ID: wpr-974568

ABSTRACT

Objective To understand the level of quality control in medical linear accelerator in Guangxi, and to provide technical support for the health administrative department to make related standards. Methods To select medical linear accelerators in Guangxi by random sampling method. Results A total of 118 accelerators were checked in 2017—2019, the initial inspection pass rates of medical linear accelerators were 56.8% and re-inspect pass rate were 96.6%. Conclusion The quality control level of the radiology institutions needs to be improved. The health administrative department should strengthen routine management,it is necessary to implement quality control in medical linear accelerator.